Eight laboratories using different liquid chromatography-mass spectrometry (LC-MS) platforms reached consistent conclusions when analyzing the same heterogeneous sample mixtures, according to a round-robin study.
The investigation examined whether untargeted metabolomic profiles could reproducibly distinguish complex samples across laboratories without relying on predefined class labels. All eight sites correctly separated the mixture types, despite detecting substantially different numbers of metabolic features.
Researchers prepared mixtures of beef, chicken, and pork extracts in known proportions, creating pure samples, two-component mixtures, and three-component mixtures at different concentrations. Each laboratory received identical aliquots of 32 samples, representing duplicate sets of 16 mixtures.
The participating laboratories used their usual reversed-phase LC-MS protocols, including Orbitrap, time-of-flight, and quadrupole time-of-flight instruments from several manufacturers. Analyses were performed in positive and negative ionization modes. This variation allowed the researchers to test whether biological patterns could be reproduced without standardizing every part of the analytical workflow.
All samples were prepared using the IROA TruQuant approach, which adds carbon-13-labeled internal standards derived from yeast. These standards supported feature filtering, quality control, and correction for ion suppression. Raw data were anonymized and processed centrally to reduce differences introduced during data handling.
Despite variation in the number of detected features (560–1,727), seven of the eight laboratories achieved perfect sample classification using non-negative matrix factorization. Suppression-corrected metabolite abundance measurements showed strong interlaboratory agreement, with a mean correlation of 0.87, increasing to 0.92 after excluding one laboratory affected by an internal-standard error.
The choice of data transformation influenced the results. Suppression correction improved direct comparisons of metabolite abundance across sites, while normalization reduced the effect of sample concentration and strengthened separation by mixture composition. However, normalization also reintroduced some laboratory-specific variation because total detected signal differed between platforms.
The researchers recommended including comprehensive standard libraries, monitoring internal standards and replicate performance, and repeating data collection when within-laboratory reproducibility is inadequate.